Setup
Mode Parameters
This section explains the parameters that can be configured on the [Sample] tab and
[Trigger] tab of the Setup dialog box, in accordance with the mode selected on the [Sample]
tab.
k Parameters Common to All Modes
The following explains the parameters that normally appear, regardless of the currently
selected mode.
Sampling Interval: Specify a value for the sampling interval. If you specify an interval of
one second, for example, a sample will be taken every second.
Number of Samples: Specify the number of samples that should be taken. The total
sampling time is calculated in accordance with the specified interval
and number of samples, and displayed on the [Single] tab of the
Sensor Setup window.
Warm-Up: Specify the warm-up time for the sensor. You can select from among
the following options for this setting.
Auto: Automatically sets a preset value for each sensor.
Manual: Selecting this option displays a dialog box like the one
shown below. Specify a warm-up time in the range of 1
to 99 seconds, and then tap [Set].
Start Trigger: Specify the event that triggers the start of sampling. You can select
from among the following options for this setting.
Tap Screen: This option specifies that sampling should start when
[OK] is tapped on the screen.
Count Down: This option specifies that sampling should start when the
end of a countdown is reached.
CH1: This option specifies that sampling should start when
CH1 channel input reaches a certain value.
SONIC: This option specifies that sampling should start when
SONIC channel input reaches a certain value.
Microphone: This option specifies that sampling should start when the
EA-200’s microphone detects a sound.