Casio FX-7400GII User Manual Page 443

  • Download
  • Add to my manuals
  • Print
  • Page
    / 483
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 442
5. Input up to 18 characters for the custom probe name, and then press E.
This will cause the highlighting to move to “Slope”.
6. Use the function keys described below to configure the custom probe setup.
To change the setting of an item, first use the f and c cursor keys to move the
highlighting to the item. Next, use the function keys to select the setting you want.
(1) Slope
Press 1(EDIT) to input the slope for the linear interpolation formula.
(2) Intercept
Press 1(EDIT) to input the intercept for the linear interpolation formula.
(3) Unit Name
Press 1(EDIT) to input up to eight characters for the unit name.
(4) Warm-up
Press 1(EDIT) to input the warm-up time.
7. Press w and then input a memory number (1 to 99).
This saves the custom probe setup and returns to the Custom Probe List, which should
now contain the new custom probe setup you configured.
u
To recall the specifications of a Vernier or CMA sensor and configure
custom probe settings
1. Perform the first two steps of the procedure under “To configure a custom probe setup”
on page 4-1.
2. Press 4(VRNR) or 5(CMA).
This displays a sensor list.
3. Use the f and c keys to move the highlighting to the sensor whose setting you want
to use as the basis of the custom probe settings, and then press w.
The name and specifications of the sensor you select will appear on the custom probe
setup screen.
To complete this procedure, perform steps 4 through 7 under “To configure a custom
probe setup” (page 4-1).
4-2
Using a Custom Probe
Page view 442
1 ... 442 443 444 ... 483

Comments to this Manuals

No comments